A-A-54072
Slit Illumination System.
The slit width shall be continuously variable from 0.0 mm to 8.0 mm, minimum.
The slit length adjustment shall be continuously variable and include the range from 1.0 mm to 8.0 mm. The slit angle shall be 90 degrees from vertical to horizontal.
Shall have aperture diaphragms with diameters of 8.0 mm, 5.0 mm, 3.0 mm, 2.0 mm, 1.0 mm and 0.2 mm.
The unit shall have at least 4 built-in filters: cobalt-blue, red-free, neutral density and heat dissipation.
Shall have a vertically adjustable chin-rest with a minimum adjustable range of 80 mm.
The optics shall be able to clearly resolve the corneal endothelium at 10X power.
Accessories.
The unit shall be supplied with the following accessories.
One fixation device.
100 chin-rest tissues.
One Goldman applanation tonometer.
One instrument table top.
Two objective lenses: lx and 1.6X.
Spare parts.
Spare illumination bulbs shall be supplied with each unit as follows:
Two fixation bulbs, two main illumination bulbs, and three of each fuse type used in the unit.
Finish.
The unit shall be fabricated from materials which are inherently resistant to corrosion or which are treated to render them corrosion-resistant. When painting or plating is used, appropriate cleaning and priming shall be performed prior to finishing. Unless otherwise specified, finish shall be equivalent or superior to that of comparable commercial equipment and shall be free of gaps, scratches, chips, and other imperfections.